Overview
WDet-5000UVI is the first full-spectrum (200-800nm) immersion UV water analyzer at home. With features of the delicate and convenient structure, plug and play, auto cleaning function, it can be widely applied to measure indexes including COD, BOD, and turbidity in surface water, underground water, and municipal sewage.
Technical Principle Of WDet-5000UVI Immersion UV Water Analyzer
The portable water quality analyser adopts a pulse xenon lamp as the light source and ultraviolet spectrometer as a detector, which can collect the absorption spectrum of water pollution between 200nm and 800nm in real time. The water pollutant content can be analyzed in real time through the chemometrics algorithm technology. The optical path is shown as the following figure.
The spectral probe can record a complete spectrum between 200nm to 800nm (UV-visible light) and distinguish the spectrum into 256 wavelengths to obtain "Spectral fingerprint". It can monitor multiple parameters simultaneously through the information contained in spectral fingerprint and compensate those parameters aiming at possible cross-sensitivity. The relevance between measuring result and laboratory result achieves high quality that simple optical instrument can never obtain before.
Features Of WDet-5000UVI Immersion UV Water Analyzer
Adopt a full spectrum (200-800nm) (UV-visible light) measurement technology and chemometrics algorithm. Compared with dual wavelength background deduction technology, it can measure more parameters and be less affected by composition change of sewage.
Use high-pressure gas to clean probe automatically, or employ immersion electrical brush to clean.
Not required for reagent preparation; save operating cost and no secondary pollution.
Fast response time(about 2s), applied to occasions with high requirements for real-time.
Without interference by chloride.
Technical Specification Of WDet-5000UVI Immersion UV Water Analyzer
Measuring Method | UV/visible light DOAS, wavelength range: 200-800nm |
Power supply | 12VDC |
Linearity | 2%F.S. |
Zero drift | <1%F.S. |
Range (COD) | 0~50~200mg/L, larger range is customizable |
Measurement interval | Minimum 60s, period can be set |
Probe immersion depth | <10m |
Optical path | Adjust according to actual working condition |
Water temp requirement | 5 ° C ~ 45 ° C |
Probe IP grade | IP68 |
Cleaning method | Gas purging or immersion electric brush |
Auto cleaning period | External control |
External interface | RS232, MODBUS protocol |
Probe dimension | Φ60mm×380.5mm |
Probe weight | < 3kg |
Auto compensation of cross sensitivity | Turbidity/organicsubstance |
Measuring range of Integrated temp | -10°C~50°C |
Temperature sensor resolution | 0.1 ° C |