🌏 Blogs / Articles / News / Classifieds / Press Releases / PSA / Podcast / Poetry / Storytelling

 

AM-5100 SO2 Analyzer

Are you looking for air analyzer equipment

Overview

AM-5100 SO2 air pollution analyzer adopts ultraviolet fluorescence technique and combines advanced micro processing technology, which could provide accurate and reliable SO2 detection at low level ppb~ppm, So we can also call it portable so2 analyzer. This analyzer can be applied to atmospheric monitoring and dilution-based CEMS system.

 

Features Of AM-5100 SO2 Analyzer

 

Measurement range 0~500ppb to 0~20ppm, user selectable;

High efficiency hydrocarbon remover and unique optical design, strong anti-interference ability;

Dark current compensation at fixed time, high stability;

Self checking function of light intensity attenuation;

Adaptive signal filtering technology to optimize response time;

Large capacity memory, automatic storage of historical data.

 

Application Of AM-5100 SO2 Analyzer

 

Atmospheric monitoring

 

Specification Of AM-5100 SO2 Analyzer

 

Range:

0~500ppb, wide range 0~20,000ppb (optional)

Unit:

ppb, ppm, ug / m3, mg / m3

Zero Point Noise:

<0.2ppb (RMS)

Low Detection Limit:

0.4ppb

Zero Drift:

<1ppb / 24h

Team Operation:

<5ppb / 24h

Response Time:

<120s (T90)

Linearity:

<1%/F.S.

Repeatability:

<1%

Sampling Flow Range:

800SCCM±10%

Ambient Temp.:

5°C~40°C

Dimension:

178mm × 430mm × 635mm

Weight:

<15kg, external pump 5kg

Power:

200V~240V, 50/60Hz, 400W

Analog Output:

10V, 5V, 1V, 0.1V (optional)

Comm.:

RS232, RS485

 8466595055?profile=original